Abstract
The aim of this study was to explore the dose response of a newly developed radio-chromic hydrogel dosimeter based on leuco malachite green dye in a gelatine matrix. The original dosimeter composition was first investigated in terms of dose response and dose-rate dependence. In addition, the initiating compounds producing chlorine radicals were substituted with compounds producing fluorine radicals, oxygen-centered radicals, carbon-centered radicals and bromine radicals. Also the surfactant was substituted by other compounds of different molecular size and charge. The original composition gave a dose response of 3.5·10-3 Gy-1cm-1 at 6 Gy/min with a dose rate dependence giving a 27 % increase when decreasing the dose rate to 1 Gy/min. None of the substituted initiating components contributed to an increase in dose response while only one surfactant increased the dose response slightly.
Cite
CITATION STYLE
Skyt, P. S., Wahlstedt, I., Yates, E. S., Muren, L. P., Petersen, J. B. B., & Balling, P. (2013). Exploring the dose response of radiochromic dosimeters. In Journal of Physics: Conference Series (Vol. 444). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/444/1/012036