Evaluation of septa quality for automatic SPME-GC-MS trace analysis

5Citations
Citations of this article
31Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

The vials used for the preparation of breath samples for automated solid-phase microextractiongas chromatographymass spectrometry analysis are crimped with septa. These septa often emit specific volatile organic compounds (VOCs) confounding the measurement results of breath samples. In the current paper, 14 different brands of magnetic caps with siliconepolytetrafluoroethylene (PTFE), butylPTFE, or butyl rubber septa were tested. The total emission of septa over a 4 h period was also evaluated. The tested septa emitted 39 different compounds, which are mainly hydrocarbons, alcohols, and ketones. Acetone and toluene are the most abundant out-gassing products. The concentration of acetone was in the range from 55 to 694 ppb for butylPTFE septum (brand 14) and butyl rubber (brand 10), respectively. The measured toluene amount was 69-1323 ppb for the septum brand 14 and brand 8 (siliconePTFE), respectively. Generally, the butyl rubber septa released higher amounts of contaminants in comparison to the silicone ones. © 2011 The Author.

Cite

CITATION STYLE

APA

Ulanowska, A., Ligor, T., Amann, A., & Buszewski, B. (2012). Evaluation of septa quality for automatic SPME-GC-MS trace analysis. Journal of Chromatographic Science, 50(1), 10–14. https://doi.org/10.1093/chromsci/bmr006