Contrast enhancement of magnetic resonance imaging (MRI) of polymer gel dosimeter by adding Platinum nano- Particles

15Citations
Citations of this article
19Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

In this research different concentrations of Platinum nanoparticles (Pt-NPs) embedded in PAGAT polymer gel and irradiated with Cobalt -60 (60Co). The OD of samples indicate significant dose enhancement caused by Pt-NPs. Different concentrations of Pt-NPs were investigated: 0.5×10-2,1×10-2 2×10-2 and 3×10-2 mg/l and these were presented the enhancement in OD by the factor of 11.56, 27.10, 15.84 and 10.07 respectively. As the mean energy of 60Co is 1.25 MeV and atomic number (Z) of Pt is 78, the predominant effect will be Compton effect and Photoelectric cannot occurred. This work studied the feasibility of using PAGAT polymer gel combined with Pt-NPs as a suitable tool to perform dosimetric investigation of nanoparticles application in radiotherapy as dosimeter and dose enhancer.

Cite

CITATION STYLE

APA

Deyhimihaghighi, N., Noor, N. M., Soltani, N., Jorfi, R., Haghir, M. E., Adenan, M. Z., … Khandaker, M. U. (2014). Contrast enhancement of magnetic resonance imaging (MRI) of polymer gel dosimeter by adding Platinum nano- Particles. In Journal of Physics: Conference Series (Vol. 546). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/546/1/012013