Abstract

In this work the results on imaging and spectroscopic performances of 14 × 14 × 1 mm CdTe detectors with 55 × 55μm and 110 × 110μm pixel pitch bump-bonded to a Timepix chip are presented. The performance of the 110 × 110μm pixel detector was evaluated at the extreme conditions beam line I15 of the Diamond Light Source. The energy of X-rays was set between 25 and 77 keV. The beam was collimated through the edge slits to 20μm FWHM incident in the middle of the pixel. The detector was operated in the time-over-threshold mode, allowing direct energy measurement. Energy in the neighbouring pixels was summed for spectra reconstruction. Energy resolution at 77 keV was found to be ΔE/E = 3.9%. Comparative imaging and energy resolution studies were carried out between two pixel size detectors with a fluorescence target X-ray tube and radioactive sources. The 110 × 110μm pixel detector exhibited systematically better energy resolution in comparison to 55 × 55μm. An imaging performance of 55 × 55μm pixellated CdTe detector was assessed using the Modulation Transfer Function (MTF) technique and compared to the larger pixel. A considerable degradation in MTF was observed for bias voltages below -300 V. Significant room for improvement of the detector performance was identified both for imaging and spectroscopy and is discussed. © 2012 IOP Publishing Ltd and SISSA.

References Powered by Scopus

Timepix, a 65k programmable pixel readout chip for arrival time, energy and/or photon counting measurements

895Citations
214Readers
Get full text

Data acquisition and processing software package for Medipix2

169Citations
37Readers
Get full text
164Citations
45Readers
Get full text

Cited by Powered by Scopus

Polarization phenomena in Al/p-CdTe/Pt X-ray detectors

37Citations
17Readers
Get full text
Get full text

This article is free to access.

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Cite

CITATION STYLE

APA

Maneuski, D., Astromskas, V., Fröjdh, E., Fröjdh, C., Gimenez, E. N., Marchal, J., … Zain, R. M. (2012). Imaging and spectroscopic performance studies of pixellated CdTe Timepix detector. Journal of Instrumentation, 7(1). https://doi.org/10.1088/1748-0221/7/01/C01038

Readers over time

‘12‘13‘14‘15‘16‘17‘18‘19‘20‘21‘22‘23‘24‘2502468

Readers' Seniority

Tooltip

PhD / Post grad / Masters / Doc 14

50%

Researcher 12

43%

Professor / Associate Prof. 2

7%

Readers' Discipline

Tooltip

Physics and Astronomy 24

86%

Engineering 3

11%

Agricultural and Biological Sciences 1

4%

Save time finding and organizing research with Mendeley

Sign up for free
0