Post-market surveillance of diagnostic test performance is critical to ensure their reliability after regulatory approval. This is especially critical in the context of the COVID-19 pandemic as the use of point-of-care tests (POCTs) became widespread. Our study focused on four rapid antigen tests (Abbott Panbio, BTNX Rapid Response, SD Biosensor, and Quidel QuickVue) and two molecular tests (Abbott ID NOW and Lucira Check IT) that were widely distributed across Canada, assessing their performance using many SARS-CoV-2 variants, including up to Omicron subvariant XBB.1.5. Overall, we found no significant difference in performance against any variant, reinforcing confidence in their use. As concerns in test efficacy have been raised by news outlets, particularly regarding the BTNX Rapid Response, this work is even more timely and crucial. Our research offers insights into the performance of widely used COVID-19 POCTs but also highlights the necessity for post-market surveillance.
Cite
CITATION STYLE
Exner, H. M., Gregorchuk, B. S. J., Castor, A.-G., Crisostomo, L., Kolsun, K., Giesbrecht, S., … Becker, M. G. (2024). Post-market surveillance of six COVID-19 point-of-care tests using pre-Omicron and Omicron SARS-CoV-2 variants. Microbiology Spectrum, 12(7). https://doi.org/10.1128/spectrum.00163-24