Abstract
The aim of this report is to address the benefits of the minimal invasive venous drainage in a pediatric cardio surgical scenario. Juvenile bovine experiments (67.4±11 kg) were performed. The right atrium was cannulated in a trans-jugular way by using the self-expandable (Smart Stat, 12/20F, 430 mm) venous cannula (Smartcannula® LLC, Lausanne, Switzerland) vs. a 14F 250 mm (Polystan Lighthouse) standard pediatric venous cannula. Establishing the cardiopulmonary bypass (CPB), the blood flows were assessed for 20 mmHg, 30 mmHg and 40 mmHg of driving pressure. Venous drainage (flow in l/min) at 20 mmHg, 30 mmHg, and 40 mmHg drainage load was 0.26±0.1, 0.35±0.2 and 0.28±0.08 for the 14F standard vs. 1.31±0.22, 1.35±0.24 and 1.9±0.2 for the Smart Stat 12/20F cannula. The 43 cm self-expanding 12/20F Smartcannula® outperforms the 14F standard cannula. The results described herein allow us to conclude that usage of the selfexpanding Smartcannula® also in the pediatric patients improves the flow and the drainage capacity, avoiding the insufficient and excessive drainage. We believe that similar results may be expected in the clinical settings. © 2010 Published by European Association for Cardio-Thoracic Surgery. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Berdajs, D. A., Bernandi, E., Burki, M., Hurni, M., Tozzi, P., Horisberger, J., & von Segesser, L. K. (2010). Self-expanding mini-cannula for remote perfusion with pediatric scenarios. Interactive Cardiovascular and Thoracic Surgery, 10(6), 873–876. https://doi.org/10.1510/icvts.2010.233080