Abstract
In macromolecular X-ray crystallography, typical data sets have substantial multiplicity. This can be used to calculate the consistency of repeated measurements and thereby assess data quality. Recently, the properties of a correlation coefficient, CC1/2, that can be used for this purpose were characterized and it was shown that CC1/2 has superior properties compared with 'merging' R values. A derived quantity, CC*, links data and model quality. Using experimental data sets, the behaviour of CC1/2 and the more conventional indicators were compared in two situations of practical importance: merging data sets from different crystals and selectively rejecting weak observations or (merged) unique reflections from a data set. In these situations controlled 'paired-refinement' tests show that even though discarding the weaker data leads to improvements in the merging R values, the refined models based on these data are of lower quality. These results show the folly of such data-filtering practices aimed at improving the merging R values. Interestingly, in all of these tests CC1/2 is the one data-quality indicator for which the behaviour accurately reflects which of the alternative data-handling strategies results in the best-quality refined model. Its properties in the presence of systematic error are documented and discussed.
Cite
CITATION STYLE
Diederichs, K., & Karplus, P. A. (2013). Better models by discarding data? Acta Crystallographica Section D: Biological Crystallography, 69(7), 1215–1222. https://doi.org/10.1107/S0907444913001121