A piezoelectric vibration-based syringe for reducing insertion force

20Citations
Citations of this article
37Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Puncturing of the human skin with a needle is perhaps the most common invasive medical procedure. Clinical studies have revealed that tissue deformation and needle deflection are the primary problem for needle misplacement in percutaneous procedures. To avoid this, various techniques for reducing insertion forces during needle insertion have been considered. This paper presents a piezoelectric vibration-based syringe to reduce insertion force. AC power was applied to the piezoelectric elements to vibrate the needle with high frequency and thereby reduce the friction and cutting forces between the needle and tissue. Vibration mode shapes of the needle were observed by finite element analysis and verified by experimental results. Effects of reducing insertion force via the vibrating needle were also confirmed by inserting the needle into the porcine tissues. The proposed syringe, which minimizes the insertion force and overcomes limitations of needle materials, can be widely utilized in robot-assisted needle insertion systems.

Cite

CITATION STYLE

APA

Huang, Y. C., Tsai, M. C., & Lin, C. H. (2012). A piezoelectric vibration-based syringe for reducing insertion force. In IOP Conference Series: Materials Science and Engineering (Vol. 42). https://doi.org/10.1088/1757-899X/42/1/012020