A simple percutaneous retrieval technique for an embolized watchman left atrial appendage closure device in the thoracic aorta using a homemade snare: A case report

3Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

A 60-year old male with paroxysmal atrial fibrillation underwent a combined procedure of left atrial appendage occlusion and pulmonary vein isolation. However, an acute intraprocedural 24-mm Watchman device dislodgement occurred, and thus a decision for urgent surgery was made. However, it was noted during the surgery that the device had migrated further to the descending aorta, just distal to the left subclavian artery. Since a right sternotomy access for retrieval was not feasible, a percutaneous approach was justified. A homemade snare was created using a combination of a long sheath, J-wire, and a regular snare, and the device was successfully retrieved without significant difficulty.

Cite

CITATION STYLE

APA

Lubis, A. C., Iqbal, M., Munawar, D. A., Hartono, B., & Munawar, M. (2021). A simple percutaneous retrieval technique for an embolized watchman left atrial appendage closure device in the thoracic aorta using a homemade snare: A case report. International Heart Journal, 62(5), 1153–1155. https://doi.org/10.1536/ihj.20-790