ProSeal Laryngeal Mask Airway (PLMA) (Teleflex Medical, Westmeath, Ireland) is a versatile and popular reusable supraglottic airway device. The manufacturer advises maximum reuse of up to 40 times. However, excessive reuse of the device can cause rare complications. Here, we report a rather uncommon complication of intraoperative cuff rupture of a PLMA probably due to the combination of a forceful metal introducer tip insertion and nitrous oxide-based anaesthesia.