A real-world study evaluating ultrasound-guided percutaneous non-targeted liver biopsy needle failures and pathology sample-quality assessment in both end-cut and side-notch needles

5Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Objectives: To determine biopsy device failures, causative factors, complications and sample quality of the 16G end-cut Biopince™ and side-notch Bard™ needles. Methods: All ultrasound-guided non-targeted liver biopsies between 01/01/2016 and 31/12/2018 were included. Operator, device, number of failures, complications and repeat biopsies were recorded. Histopathology samples were reviewed for all cases of needle failure and a group with no failures, and graded "yes/no" for the presence of steatosis, inflammation and fibrosis. The pathology slides from these cases were reviewed to assess biopsy sample quality (length and portal tract number). The failure and no-failure groups were compared in terms of device type/histology, and sample quality was compared between the needle types. Results: 1004 patients were included. 93.8% (n = 942) required one needle pass to obtain a sample and 6.2% (n = 62) required >1 pass due to needle failure. Total of 76 needle failures, more with end-cut than side-notch needles (8.7% vs 2.9%) (p < 0.001). No needle failures resulted in complication. The presence of liver fibrosis was associated with fewer needle failures (p = 0.036). The major complication rate was 0.4% (4/1044). A biopsy with >10 portal tracts was obtained in 90.2% of specimens > 20 mm long, compared with 66% of 16-20 mm biopsies and 21% of <16 mm biopsies. The target of >10 portal tracts was achieved in 10/26 (38.5%) of side-notch biopsies and 64/90 (71.1%) of end-cut biopsies (p = 0.004). Conclusion: Ultrasound-guided liver biopsy is safe and sample quality is consistently good when a core >20 mm long is obtained. The end-cut biopsy device generated reliably good quality biopsy samples; however, the needle failure rate was significantly higher than the side-cut needle.

Cite

CITATION STYLE

APA

ZHONG, J., ALLARD, R., HEWITSON, D., WESTON, M., HULSON, O., BURBIDGE, S., … ALBAZAZ, R. (2021). A real-world study evaluating ultrasound-guided percutaneous non-targeted liver biopsy needle failures and pathology sample-quality assessment in both end-cut and side-notch needles. British Journal of Radiology, 94(1125). https://doi.org/10.1259/bjr.20210475

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free