A high-order harmonic generation apparatus for time- and angle-resolved photoelectron spectroscopy

83Citations
Citations of this article
105Readers
Mendeley users who have this article in their library.

Abstract

We present a table top setup for time- and angle-resolved photoelectron spectroscopy to investigate band structure dynamics of correlated materials driven far from equilibrium by femtosecond laser pulse excitation. With the electron-phonon equilibration time being in the order of 1-2 ps it is necessary to achieve sub-picosecond time resolution. Few techniques provide both the necessary time and energy resolution to map non-equilibrium states of the band structure. Laser-driven high-order harmonic generation is such a technique. In our experiment, a grating monochromator delivers tunable photon energies up to 40 eV. A photon energy bandwidth of 150 meV and a pulse duration of 100 fs FWHM allow us to cover the k-space necessary to map valence bands at different k z and detect outer core states. © 2013 AIP Publishing LLC.

Cite

CITATION STYLE

APA

Frietsch, B., Carley, R., Döbrich, K., Gahl, C., Teichmann, M., Schwarzkopf, O., … Weinelt, M. (2013). A high-order harmonic generation apparatus for time- and angle-resolved photoelectron spectroscopy. Review of Scientific Instruments, 84(7). https://doi.org/10.1063/1.4812992