Unpredictability of delivered bubble nasal continuous positive airway pressure: Role of bias flow magnitude and nares-prong air leaks

56Citations
Citations of this article
48Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Nasal continuous-positive-airway-pressure (NCPAP) is popular for infant respiratory support. We compared delivered to intended intra-prong, proximal-airway, and distal-airway pressures using ventilator (V-NCPAP) and bubble (B-NCPAP) devices. Measurements were repeated at five flows (4, 6, 8, 10, and 12 L/min) and three NCPAP (4, 6, and 8 cm H2O) under no, small, and large nares-prong interface leak conditions. With no-leak, delivered B-NCPAP was systematically greater than intended levels at all pressure sites. The corresponding V-NCPAP flow-dependence was none-to-minimal. Prong and intra-airway B-NCPAP overshoots were also observed with small-leak, while only prong B-NCPAP showed a flow-dependent overshoot for large-leak. Leaks did not affect intra-prong V-NCPAP but resulted in progressively lower than desired, flow-independent intra-airway V-NCPAP. We conclude that the self-adjusting capability of ventilators allows closely matched actual versus intended V-NCPAP. Alternatively, for the range of flows used clinically, intra-prong and intra-airway B-NCPAP are systematically higher at increasing flows than operator-intended levels, even when appreciable nares-prong leak is present. Additionally, the oscillations (noise) characterizing B-NCPAP are substantially attenuated between the proximal and distal airways; therefore, it is unlikely that B-NCPAP engenders ventilation or lung recruitment via this phenomenon. Tubing submersion depth for setting the level of B-NCPAP is highly inaccurate, and operators should instead rely on intra-prong pressure measurement. © International Pediatrics Research Foundation, Inc. 2007. All Rights Reserved.

Cite

CITATION STYLE

APA

Kahn, D. J., Courtney, S. E., Steele, A. M., & Habib, R. H. (2007). Unpredictability of delivered bubble nasal continuous positive airway pressure: Role of bias flow magnitude and nares-prong air leaks. Pediatric Research, 62(3), 343–347. https://doi.org/10.1203/PDR.0b013e318123f702

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free