Corresponding dose response of radiographic film with layered Gafchromic film

30Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This note investigates the dose response of layered HS Gafchromic film compared to Kodak EDR-2 radiographic film. Using five layers of HS type Gafchromic film a dose response greater than EDR-2 film is achieved at the peak wavelenth (0.55 OD/Gy versus 0.3 OD/Dy for EDR-2 film). Even over a broader waveband of 30 nm, which is similar to that found in ultra bright LED scanners, the response was found to be 0.38 OD/Gy as opposed to 0.29 OD/Gy. Measurements averaged over the entire visible spectrum produce a relative dose response of 0.165 OD/Gy for five layer HS and 0.29 OD/Gy for EDR-2 film. Due to this high dose response that is achievable, the five layer HS could be used in applications where small doses are delivered to certain areas and a low dependence of energy response is required for measurement.