Abstract
Conventional X-ray is commonly used for pelvic fracture follow-ups, but has a precision of only ± 5 mm. Implantation of tantalum beads together with RSA has shown high precision but not applicable in clinical practice. CT scan has been shown a suitable substitute for RSA to follow the metal markers. We aimed to assess whether implantation of metal markers could be avoided using CT scan and merging of bone surface anatomy. A human cadaveric pelvis marked with 0.8 mm tantalum beads was fixed over the symphysis and the right SI-joint. Left hemi-pelvis was subsequently distracted using plastic spacers. Sequential CT exams was conducted and data were analyzed using Sectra ® (Sectra AB), CTMA package. Examinations were repeated after 2 weeks. Bone registration showed better precision than registration based on tantalum beads. However, only the difference in angular changes was statistically significant (p = 0.008). The confidence interval of the repeatability was ± 0.5 mm for translation and ± 0.5° for rotation. This new non-invasive technique showed good precision and repeatability and might be a future option in clinical practice for post-operative follow-ups of patients with pelvic fractures.
Author supplied keywords
Cite
CITATION STYLE
Bakhshayesh, P., Olivecrona, H., Weidenhielm, L., & Enocson, A. (2019). A New CT Based Method for Post-operative Motion Analysis of Pelvic Fractures. 3D Research, 10(1). https://doi.org/10.1007/s13319-019-0217-z