Abstract
This study presents the optimization and validation of methods for the analysis of retinol, thiamine, niacin, pyridoxine, folic acid, cyanocobalamin, zinc, and iron in fortified kernels (coated and extruded) and in fortified rice. The analyses were performed by HPLC-UV/FLD/MS and ICP-OES. The optimized methods showed good resolution of the analyte peaks, excellent recovery (87-108%), reproducibility with relative standard deviation (SD) of analyte content between 1.8 and 11% and high correlation coefficient of the calibration curves (R2 > 0.997). Limit of detection was from 2.8 E-4 mg/kg for pyridoxine to 1.26 mg/kg for zinc and limit of quantification was from 9.2 E-4 mg/kg for pyridoxine to 4.21 mg/kg for zinc. Thereby the optimized methods demonstrated reliability and sensitivity in the detection and quantification of these micronutrients and that they are suitable for routine analysis of fortified kernels (coated and extruded) and fortified rice.
Cite
CITATION STYLE
Ivarsen, E., Andersen, C. P., Jensen, S. M., Pedersen, C. T., & Svaneborg, A. K. (2021). Quantification of retinyl palmitate, thiamine, niacin, pyridoxine, folic acid, cyanocobalamin, zinc, and iron by chromatographic methods in fortified kernels and fortified rice. Acta Chromatographica, 33(2), 145–152. https://doi.org/10.1556/1326.2020.00770