Abstract
We have developed a high sensitivity time-resolved two-photon scanning microscope. At an excitation wavelength of 960 nm, a spatial point spread function of 0.3 μm (FWHM) radially and 0.9 μm (FWHM) axially is measured for an 1.25 N.A. objective. The light source is a mode-locked titanium-sapphire laser. The time resolution is 400 ps with common chromophores used in microscopy. Time resolution is obtained using the frequency-domain heterodyning technique in which the laser is synchronized at a very high cross-correlation frequency to the rest of the electronics. We demonstrate spatial and time resolution using well-characterized fluorescent microspheres. We show two applications of two-photon time-resolved fluorescence microscopy: time-resolved imaging of multiple dye labeled cells and quantitative cellular calcium concentration using a lifetime indicator.
Author supplied keywords
Cite
CITATION STYLE
So, P. T. C., French, T., Yu, W. M., Berland, K. M., Dong, C. Y., & Gratton, E. (1995). Time-resolved fluorescence microscopy using two-photon excitation. Bioimaging, 3(2), 49–63. https://doi.org/10.1002/1361-6374(199506)3:2<49::AID-BIO1>3.0.CO;2-E