Frequency response of PVDF needle-type hydrophones

46Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.
Get full text

Abstract

This paper examines the factors governing the frequency response of ultrasonic polyvinylidene flouride (PVDF) polymer needle-type hydrophones, in particular the sensitivity variations in the lower frequency range of 1-6 MHz. A theoretical model was used to analyze the influence of the hydrophone's diameter, the metal electrodes, thickness, PVDF material properties, the adhesive layer acoustical characteristics and the backing material, on the frequency response of the hydrophone. The results of the theoretical modelling differ by less than ± 0.5 dB from those obtained experimentally from the reciprocity calibration in the frequency range 1-20 MHz. It is shown that the needle hydrophone's diameter and backing material are the main reasons for the sensitivity variations observed in the frequency range below 6 MHz. © 1994.

Cite

CITATION STYLE

APA

Fay, B., Ludwig, G., Lankjaer, C., & Lewin, P. A. (1994). Frequency response of PVDF needle-type hydrophones. Ultrasound in Medicine and Biology, 20(4), 361–366. https://doi.org/10.1016/0301-5629(94)90004-3