Vibration as a pitfall in pyrosequencing analyses

3Citations
Citations of this article
13Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Since methylation analysis has become an important tool in forensic genetics, the reliability and credibility of the method must be ensured. After a successful validation and establishment of several pyrosequencing assays using a PyroMark® Q48 Autoprep instrument (Qiagen, Hilden, Germany), we decided to expand the method further purchasing a second instrument. But after initializing this second instrument side by side with the first, the majority of analyses failed (97 samples of 133 samples (73%)). The number of error messages increased rapidly and the average RFU values decreased. After purchasing two anti-vibration weighing tables for the PyroMark® instruments and repeating the analyses under the same conditions and with identical samples the results improved considerably, 115 samples of 130 samples (88%) showed successful and reproducible results. These findings demonstrate the impact of vibrations and percussions on PyroMark® Q48 Autoprep performance and the reliability of methylation analyses.

Cite

CITATION STYLE

APA

Konrad, H., Schäfer, L., Sturm, H., Hördt, L., Bajanowski, T., & Poetsch, M. (2022). Vibration as a pitfall in pyrosequencing analyses. International Journal of Legal Medicine, 136(1), 103–105. https://doi.org/10.1007/s00414-021-02716-7