Abstract

Purpose To study the accommodation process in normal eyes using a commercially available clinical system based on swept-source anterior segment optical coherence tomography (AS-OCT). Setting Ophthalmology Department, University of Parma, Italy. Design Evaluation of diagnostic technology. Methods Right eyes were analyzed using swept-source AS-OCT (Casia SS-1000). The optical vergence of the internal coaxial fixation target was adjusted during imaging to obtain monocular accommodation stimuli with different amplitudes (0, 3.0, 6.0, and 9.0 diopters [D]). Overlapping of real and conjugate OCT images enabled imaging of all the anterior segment optical surfaces in a single frame. Central corneal thickness (CCT), anterior chamber depth (ACD), and lens thickness were extracted from the OCT scans acquired at different static accommodation stimulus amplitudes. The crystalline lens was analyzed dynamically during accommodation and disaccommodation by acquiring sequential OCT images of the anterior segment at a rate of 8 frames per second. The lens thickness was extracted from the temporal sequence of OCT images and plotted as a function of time. Results The study analyzed 14 eyes of 14 subjects aged 18 to 46 years. During accommodation, the decrease in the ACD was statistically significant (P

Cite

CITATION STYLE

APA

Neri, A., Ruggeri, M., Protti, A., Leaci, R., Gandolfi, S. A., & Macaluso, C. (2015). Dynamic imaging of accommodation by swept-source anterior segment optical coherence tomography. Journal of Cataract and Refractive Surgery, 41(3), 501–510. https://doi.org/10.1016/j.jcrs.2014.09.034

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free