Abstract
Despite technological advances in a newer generation of ventricular assist devices (VAD), complications, such as pump thromboses, remain a significant cause of morbidity and indeed mortality in these patients. We present the case of a 34-year old patient who underwent HeartAssist 5 (HA5) implantation as a bridge to cardiac transplant. After an initial uneventful recovery, he developed a pump thrombosis that was refractory to medical treatment. We present the surgical technique used to exchange the HA5 with a HeartWare (HVAD), leaving the old inflow-sewing ring in situ. © The Author 2013. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Sáez, D. G., Mohite, P., Zych, B., Sabashnikov, A., & Simon, A. R. (2013). Successful replacement of a HeartAssist 5 ventricular assist device with a HeartWare without removal of the original sewing/attachment rings: How to do it. Interactive Cardiovascular and Thoracic Surgery, 16(6), 888–889. https://doi.org/10.1093/icvts/ivt066
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.