Abstract
The effectiveness of an adaptive directional microphone design, as implemented in the Phonak Claro behind-the-ear hearing aid, is evaluated. Participants were fit bilaterally and tested in 2 environments, an anechoic chamber and a moderately reverberant classroom, with the microphones in the fixed (cardioid) setting and the adaptive setting. Five speakers were placed between 110° and 250° azimuth around the listener. Speech-weighted noise was presented from those speakers at an overall level (OAL) of 65 dB (A). Noise was increased by 8 dB from 1 speaker at a time, using 2-s modulation and random assignment, while the output from the other speakers was reduced to maintain the constant OAL. Results of 2 speech perception tasks used as outcome measures indicated that the adaptive system was not able to follow the dominant noise source in the presence of lower level noise sources. Self-report measures obtained after blinded home trials were consistent with laboratory findings that the participants did not perceive this adaptive microphone design to be more effective than the default fixed-microphone option. © American Speech-Language-Hearing Association.
Author supplied keywords
Cite
CITATION STYLE
Bentler, R. A., Tubbs, J. L., Egge, J. L. M., Flamme, G. A., & Dittberner, A. B. (2004). Evaluation of an adaptive directional system in a DSP hearing aid. American Journal of Audiology, 13(1), 73–79. https://doi.org/10.1044/1059-0889(2004/010)