Abstract
The Directigen Flu A+B enzyme immunoassay and the Abbott TestPack RSV enzyme immunoassay were each compared with a multiplex RT-PCR ELISA by testing 635 nasopharyngeal aspirates collected from children aged < 16 years who had been hospitalised with acute respiratory tract infection during the epidemic season 2002-2003. In this study, the sensitivity of the Directigen Flu A+B assay was unacceptably low (29.3% and 10.0%, respectively) for the detection of influenza A and B viruses. The sensitivity of the Abbott TestPack RSV assay (77.4%) was acceptable and in agreement with the multiplex RT-PCR ELISA. © 2005 Copyright by the European Society of Clinical Microbiology and Infectious Diseases.
Author supplied keywords
Cite
CITATION STYLE
Gröndahl, B., Puppe, W., Weigl, J., & Schmitt, H. J. (2005). Comparison of the BD Directigen Flu A+B Kit and the Abbott TestPack RSV with a multiplex RT-PCR ELISA for rapid detection of influenza viruses and respiratory syncytial virus. Clinical Microbiology and Infection, 11(10), 848–850. https://doi.org/10.1111/j.1469-0691.2005.01223.x