Abstract
Background: Magnetic resonance imaging (MRI) has unique advantages for guiding interventions, but the narrow space is a major challenge. This study evaluates the feasibility of a remote-controlled hydrostatic actuator system for MRI-guided targeted needle placement. Methods: The effects of the hydrostatic actuator system on MR image quality were evaluated. Using a reference step-and-shoot method (SS) and the proposed actuator-assisted method (AA), two operators performed MRI-guided needle placement in targets (n = 12) in a motion phantom. Results: The hydrostatic actuator system exhibited negligible impact on MR image quality. In dynamic targets, AA was significantly more accurate and precise than SS, with mean ± SD needle-to-target error of 1.8 ± 1.0 mm (operator 1) and 1.3 ± 0.5 mm (operator 2). AA reduced the insertion time by 50% to 80% and total procedure time by 25%, compared to SS. Conclusions: The proposed hydrostatic actuator system may improve accuracy and reduce procedure time for MRI-guided targeted needle placement during motion.
Author supplied keywords
Cite
CITATION STYLE
Mikaiel, S., Simonelli, J., Li, X., Lee, Y. H., Lee, Y. S., Sung, K., … Wu, H. H. (2020). MRI-guided targeted needle placement during motion using hydrostatic actuators. International Journal of Medical Robotics and Computer Assisted Surgery, 16(2). https://doi.org/10.1002/rcs.2041