Summary We measured heating of isotonic saline by three fluid warmers in six experiments: saline at 5 C or 20 C delivered at 30, 50 or 100 ml.min-1. At the three flow rates, the enFLOW®, buddy lite™ and ThermoSens® systems heated 5 C saline to mean (SD) temperatures of: 41.1 (0.5) C, 37.7 (0.6) C and 39.1 (0.6) C; to 40.3 (0.8) C, 33.9 (1.6) C and 39.3 (0.7) C; and to 37.1 (0.8) C, 24.0 (1.3) C and 37.6 (1.0) C, respectively, p < 0.0001 for each experiment. The mean (SD) times taken to heat 5 C saline were: 16.6 (1.7) s, 258.4 (58.9) s and 134.2 (79.6) s; 16.9 (1.8) s, 256.2 (62.2) s and 182.5 (74.5) s; and 21.5 (1.5) s, 275.9 (49.3) s and 313.5 (18.0) s, respectively, p < 0.0003 for each experiment. The results for saline at 20 C were similar. The enFLOW system heated saline above 36 C faster than the ThermoSens system, whereas the buddy lite often failed to achieve 36 C.
CITATION STYLE
Kim, H. J., Yoo, S. M., Jung, J. S., Lee, S. H., Sun, K., & Son, H. S. (2015). The laboratory performance of the enFLOW®, buddy liteTM and ThermoSens® fluid warmers. Anaesthesia, 70(2), 205–208. https://doi.org/10.1111/anae.12906