Abstract
Duty cycle-based trapping and extraction processes have been investigated for linear digitally driven multipoles by simulating ion trajectories. The duty cycles of the applied waveforms were adjusted so that an effective trapping or ejection electric field was created between the rods and the grounded end cap electrodes. By manipulating the duty cycles of the waveforms, the potentials of the multipole rods can be set equal for part of the waveform cycle. When all rods are negative for this period, the device traps positive ions and when all are positive, it ejects them in focused trajectories. Four Linac II electrodes [1] have been added between the quadrupole rods along the asymptotes to create an electric field along the symmetry axis for collecting the ions near the exit end cap electrode and prompt ejection. This method permits the ions to be collected and then ejected in a concentrated and collimated plug into the acceleration region of a time-of-flight mass spectrometer (TOFMS). Our method has been shown to be independent of mass. Because the resolution of orthogonal acceleration TOFMS depends primarily on the dispersion of the ions injected into the acceleration region and not on the ion mass, this technology will enable high resolution in the ultrahigh mass range (m/z > 20,000). © 2011 Elsevier B.V. All rights reserved.
Author supplied keywords
Cite
CITATION STYLE
Lee, J., Marino, M. A., Koizumi, H., & Reilly, P. T. A. (2011). Simulation of duty cycle-based trapping and ejection of massive ions using linear digital quadrupoles: The enabling technology for high resolution time-of-flight mass spectrometry in the ultra high mass range. International Journal of Mass Spectrometry, 304(1), 36–40. https://doi.org/10.1016/j.ijms.2011.03.011
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.