Comparison of blood ΒHBA measurement devices for diagnosis of subclinical pregnancy toxaemia in sheep: A field study

7Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Pregnancy toxaemia is a serious disease that occurs during the last trimester of pregnancy in sheep. Yet, in most cases, the disease may have a subclinical course. This study was aimed at comparing blood βHBA measurement devices for diagnosis of subclinical pregnancy toxaemia in late pregnant sheep. In the study, a total of 50 blood samples were collected from Romanov (n = 30) and cross-bred Hamdani (n = 20) sheep. Blood βHBA levels were measured using TaiDoc TD-4235 and CentriVet βHBA hand-held meter. Randox βHBA (Ranbut) assay was used as a reference laboratory method to compare hand-held meters. βHBA value of 0.8 mmol/L was set as the cut-off value for diagnosis of subclinical pregnancy toxaemia. Statistical analyses were carried out using Minitab 21 and Jamovi software. In the study, the correlation of Randox-TaiDoc TD-4235 and Randox-CentriVet was.822 (p

Cite

CITATION STYLE

APA

Turgut, A. O., Koca, D., & Ünver, A. (2024). Comparison of blood ΒHBA measurement devices for diagnosis of subclinical pregnancy toxaemia in sheep: A field study. Reproduction in Domestic Animals, 59(5). https://doi.org/10.1111/rda.14589