Calibration of gradient propagation delays for accurate two-dimensional radiofrequency pulses

13Citations
Citations of this article
25Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Hardware-related delays between the requested and actual start times of the gradient waveforms on each physical axis are of particular importance for multidimensional selective excitation in which the synchronization of gradient and radiofrequency (RF) waveforms is critical. A method is proposed for the accurate calibration of gradient propagation delays to optimize the spatial accuracy of 2D RF pulses, although the results may also be used to reduce artifacts in other MR techniques. The sensitivity of 20 RF pulses to unconnected time shifts between the gradient and RF waveforms was exploited to calibrate accurately the propagation delays on each physical gradient axis. This was achieved using a technique that relates the effect of gradient delays in the component waveforms of a constant-angular rate spiral k-space trajectory 2D RF pulse to the spatial location of the subsequent excitation profile. Comparison was also made with a procedure based on a previously described k-space plotting method, showing broad agreement, but with some discrepancies that illustrate the value of a self-referenced correction method for multidimensional RF pulses. © 2004 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Davies, N. P., & Jezzard, P. (2005). Calibration of gradient propagation delays for accurate two-dimensional radiofrequency pulses. Magnetic Resonance in Medicine, 53(1), 231–236. https://doi.org/10.1002/mrm.20308

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free