Background: Outside of a clinical setting, oscillometric devices make remote monitoring of blood pressure and virtual care more convenient and feasible. HeartBeat Technologies Ltd developed a novel approach to measuring blood pressure remotely after an initial blood pressure reading by a nurse using the conventional measurement method. Using a finger pulse oximeter, a photoplethysmogram wave is transmitted by Bluetooth to a smartphone or tablet. A smartphone app (MediBeat) transmits the photoplethysmogram to a server for analysis by a proprietary algorithm-the person's current blood pressure is sent back to the smartphone and to the individual's health care provider. Objective: This study sought to determine whether the HeartBeat algorithm calculates blood pressure as accurately as required by the European Society of Hypertension International Protocol revision 2010 (ESH-IP2) for validation of blood pressure measuring devices. Methods: ESH-IP2 requirements, modified to conform to a more recent international consensus statement, were followed. The ESH-IP2 establishes strict guidelines for the conduct and reporting of any validation of any device to measure blood pressure, including using the standard manual blood pressure instrument as a comparator and specific required accuracy levels for low, medium, and high ranges of blood pressure readings. The consensus statement requires a greater number of study participants for each of the blood pressure ranges. The validation of the accuracy of the algorithm was conducted with a Contec CMS50EW pulse oximeter and a Samsung Galaxy XCover 4 smartphone. Results: The differences between the HeartBeat-calculated and the manually measured blood pressures of 62 study participants did not meet the ESH-IP2 standards for accuracy for either systolic or diastolic blood pressure measurements. There was no discernible pattern in the inaccuracies of the HeartBeat-calculated measurements. Conclusions: The October 4, 2019 version of the HeartBeat algorithm, implemented in combination with the MediBeat app, a pulse oximeter, and an Android smartphone, was not sufficiently accurate for use in a general adult population.
CITATION STYLE
Holyoke, P., Yogaratnam, K., & Kalles, E. (2021). Web-based smartphone algorithm for calculating blood pressure from photoplethysmography remotely in a general adult population: Validation study. Journal of Medical Internet Research, 23(4). https://doi.org/10.2196/19187
Mendeley helps you to discover research relevant for your work.