Abstract
The aim of this study is to develop polymeric chemisorbents with a high density of ninhydrin groups, able to covalently bind urea under physiological conditions and thus potentially suitable for use in a wearable artificial kidney. Macroporous beads are prepared by suspension polymerization of 5-vinyl-1-indanone (vinylindanone) using a 90:10 (v/v) mixture of toluene and nitrobenzene as a porogen. The indanone groups are subsequently oxidized in a one-step procedure into ninhydrin groups. Their urea absorption kinetics are evaluated under both static and dynamic conditions at 37 °C in simulated dialysate (urea in phosphate buffered saline). Under static conditions and at a 1:1 molar ratio of ninhydrin: urea the sorbent beads remove ≈0.6–0.7 mmol g−1 and under dynamic conditions and at a 2:1 molar excess of ninhydrin ≈0.6 mmol urea g−1 sorbent in 8 h at 37 °C, which is a step toward a wearable artificial kidney.
Author supplied keywords
Cite
CITATION STYLE
Jong, J. A. W., Guo, Y., Hazenbrink, D., Douka, S., Verdijk, D., van der Zwan, J., … van Nostrum, C. F. (2020). A Ninhydrin-Type Urea Sorbent for the Development of a Wearable Artificial Kidney. Macromolecular Bioscience, 20(3). https://doi.org/10.1002/mabi.201900396
Register to see more suggestions