Pressures within air-filled tracheal cuffs at altitude - An in vivo study

22Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Aeromedical transport of mechanically ventilated critically ill patients is now a frequent occurrence. However, the performance of the air filled tracheal tube cuff at altitude has not been studied in vivo. We measured the tracheal cuff pressures at ground level and at 3000 ft, in 10 intubated patients. With air providing the seal in the cuff the mean rise in cuff pressure was 23 cmH2O, which took the pressures above the critical perfusion pressure of the tracheal mucosa. This could lead to tracheal injury. © 2004 Blackwell Publishing Ltd.

Cite

CITATION STYLE

APA

Henning, J., Sharley, P., & Young, R. (2004). Pressures within air-filled tracheal cuffs at altitude - An in vivo study. Anaesthesia, 59(3), 252–254. https://doi.org/10.1111/j.1365-2044.2004.03609.x