Comparison of global end-diastolic volume index derived from jugular and femoral indicator injection: a prospective observational study in patients equipped with both a PiCCO-2 and an EV-1000-device

4Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Transpulmonary thermodilution (TPTD)-derived global end-diastolic volume index (GEDVI) is a static marker of preload which better predicted volume responsiveness compared to filling pressures in several studies. GEDVI can be generated with at least two devices: PiCCO and EV-1000. Several studies showed that uncorrected indicator injection into a femoral central venous catheter (CVC) results in a significant overestimation of GEDVI by the PiCCO-device. Therefore, the most recent PiCCO-algorithm corrects for femoral indicator injection. However, there are no systematic data on the impact of femoral indicator injection for the EV-1000 device. Furthermore, the correction algorithm of the PiCCO is poorly validated. Therefore, we prospectively analyzed 14 datasets from 10 patients with TPTD-monitoring undergoing central venous catheter (CVC)- and arterial line exchange. PiCCO was replaced by EV-1000, femoral CVCs were replaced by jugular/subclavian CVCs and vice-versa. For PiCCO, jugular and femoral indicator injection derived GEDVI was comparable when the correct information about femoral catheter site was given (p = 0.251). By contrast, GEDVI derived from femoral indicator injection using the EV-1000 was obviously not corrected and was substantially higher than jugular GEDVI measured by the EV-1000 (846 ± 250 vs. 712 ± 227 ml/m2; p = 0.001). Furthermore, measurements of GEDVI were not comparable between PiCCO and EV-1000 even in case of jugular indicator injection (p = 0.003). This is most probably due to different indexations of the raw value GEDV. EV-1000 could not be recommended to measure GEDVI in case of a femoral CVC. Furthermore, different indexations used by EV-1000 and PiCCO should be considered even in case of a jugular CVC when comparing GEDVI derived from PiCCO and EV-1000.

Cite

CITATION STYLE

APA

Herner, A., Heilmaier, M., Mayr, U., Schmid, R. M., & Huber, W. (2020). Comparison of global end-diastolic volume index derived from jugular and femoral indicator injection: a prospective observational study in patients equipped with both a PiCCO-2 and an EV-1000-device. Scientific Reports, 10(1). https://doi.org/10.1038/s41598-020-76286-w

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free