Online blind calibration of non-uniform photodetectors: Application to endomicroscopy

N/ACitations
Citations of this article
15Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

We present an original method for the online blind calibration of non-uniform photodetectors. The disparity of the detectors may arise from both irregular spatial arrangement and distinct slowly time-varying photometric transfer functions. As natural images are mostly continuous, the signal collected by neighboring detectors is strongly correlated over time. The core idea of our method is to translate the calibration problem into relative pairwise calibrations between neighboring detectors followed by the regularized inversion of a system akin to gradient-based surface recovery. From our blind calibration procedure, we design an online blind calibration pipeline compatible with clinical practice. Online blind calibration is proved to be statistically better than standard offline calibration for reconstructing endomicroscopy sequences.

Cite

CITATION STYLE

APA

Savoire, N., André, B., & Vercauteren, T. (2012). Online blind calibration of non-uniform photodetectors: Application to endomicroscopy. In Lecture Notes in Computer Science (including subseries Lecture Notes in Artificial Intelligence and Lecture Notes in Bioinformatics) (Vol. 7512 LNCS, pp. 639–646). Springer Verlag. https://doi.org/10.1007/978-3-642-33454-2_79