Abstract
This study evaluated the feasibility of utilizing a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance (QA) in intensity-modulated radiotherapy (IMRT). Contoured left and right head phantoms were converted from DICOM to STL format. Fused deposition modeling (FDM) was used to construct an anthropomorphic patient-specific head phantom with a 3D printer. An established QA technique and the patient-specific head phantom were used to compare the calculated and measured doses. When the established technique was used to compare the calculated and measured doses, the gamma passing rate for γ ≤ 1 was 97.28%, while the gamma failure rate for γ > 1 was 2.72%. When the 3D-printed patient-specific head phantom was used, the gamma passing rate for γ ≤ 1 was 95.97%, and the gamma failure rate for γ > 1 was 4.03%. The 3D printed patient-specific head phantom was concluded to be highly feasible for patient-specific QA prior to complicated radiotherapy procedures such as IMRT.
Cite
CITATION STYLE
Yea, J. W., Park, J. W., Kim, S. K., Kim, D. Y., Kim, J. G., Seo, C. Y., … Oh, S. A. (2017). Feasibility of a 3D-printed anthropomorphic patient-specific head phantom for patient-specific quality assurance of intensity-modulated radiotherapy. PLoS ONE, 12(7). https://doi.org/10.1371/journal.pone.0181560