Enzyme Biosensor for Detection of Organophosphate Pesticide Residues Base on Screen Printed Carbon Electrode (SPCE)-Bovine Serum Albumin (BSA)

  • Mulyasuryani A
  • Dofir M
N/ACitations
Citations of this article
22Readers
Mendeley users who have this article in their library.

Abstract

The maximum level of organophosphate pesticide residues in rice is 0.1 mg/kg and 0.5 mg/kg in vegetables. The control of pesticide residues in agricultural products required a method of analysis quickly and accurately. The research developed a biosensor for the detection of organophosphate pesticide residues in agricultural products. The research studied immobilized organophosphate hydrolase (OPH) mass and characterization of biosensor. The solution conductivity measurement in the conductivity cell consists of a 1 × 5 mm2 pair of electrodes screen printed carbon electrode (SPCE). The instrument is a converted local conductometer. From the results of study concluded that the optimum performance of the biosensor was obtained from the 105 µg OPH, at pH 8.5 with a response time of 45 seconds. In that condition the sensitivity of biosensor is 28.04 μS/ppm and 0.18 ppm detection limit and the maximum concentration of pesticide which can be measured is 1 ppm.

Cite

CITATION STYLE

APA

Mulyasuryani, A., & Dofir, M. (2014). Enzyme Biosensor for Detection of Organophosphate Pesticide Residues Base on Screen Printed Carbon Electrode (SPCE)-Bovine Serum Albumin (BSA). Engineering, 06(05), 230–235. https://doi.org/10.4236/eng.2014.65027

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free