Abstract

Background: Because standard home oxygen (O2) systems deliver O2 at fixed rates, these systems are not designed to ensure optimal oxygen delivery based on physiologic need. We tested the ability of the AccuO2 (OptiSat Medical, Minneapolis, Minnesota), a portable, closed-loop, oximetrydriven, O2-conserving device to maintain SpO2 at ≥ 90%, compared to continuous-flow oxygen and a standard O2-conserving device (CR-50, Puritan-Bennett, Pleasanton, California). Methods: We randomly assigned 28 patients who were on continuous home O2 for COPD to use each of 3 O2 delivery systems (continuous-flow O2, CR-50, and AccuO2) for 8 hours a day, for 2 consecutive days, at home, at their current O2 prescription. We recorded SpO2 and calculated the conservation ratio (duration of a given O2 supply with an O2-conserving device compared to continuous-flow O2). Results: Twenty-two patients completed all 3 study arms; 2 additional patients completed the AccuO2 arm and the continuous-flow O2 arm. The mean ± SD SpO2 was 92 ± 4% with continuousflow O2, 92 ± 4% with the CR-50, and 91 ± 2% with AccuO2 (P =.006 for the AccuO2 vs continuous-flow O2, P =.03 for the AccuO2 vs the CR-50). SpO2 variability was less with the AccuO2 (P

Cite

CITATION STYLE

APA

Rice, K. L., Schmidt, M. F., Buan, J. S., Lebahn, F., & Schwarzock, T. K. (2011). AccuO2 oximetry-driven oxygen-conserving device versus fixed-dose oxygen devices in stable COPD patients. Respiratory Care, 56(12), 1901–1905. https://doi.org/10.4187/respcare.01059

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free