The aim of this study is to assess the clinical performance of 6 different low-flux dialysis membranes under steady-state conditions in terms of urea and phosphate clearances. Ten stable hemodialysis patients were examined. The following dialyzers were studied, all in 1.5to 1.6-m2 format: cuprammonium, cellulose acetate, cellulose diacetate, hemophane, polysulfone (low-flux), and polysynthane. The following parameters were examined: urea reduction ratio, phosphate reduction ratio, 'instantaneous dialyzer clearance' for urea and phosphate, and total amount of urea and phosphate removed in the dialysate over a 1-week (three dialyses) period. Although there were differences between the membranes, all produced results within a narrow range. There was no one membrane that produced superior clearances in all categories. The cellulose acetate membrane was the least satisfactory membrane. Phosphate clearances were at best one third that of urea clearances. When choosing a low-flux dialysis membrane, urea and phosphate clearances are so similar amongst different membranes that other criteria are likely to have a greater influence on the choice of membrane.

Cite

CITATION STYLE

APA

Kerr, P. G., Lo, A., Chin, M. M., & Atkins, R. C. (1999). Dialyzer performance in the clinic: Comparison of six low-flux membranes. Artificial Organs, 23(9), 817–821. https://doi.org/10.1046/j.1525-1594.1999.06297.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free