Electrochemical detection of creatinine: exploiting copper(ii) complexes at Pt microelectrode arrays

25Citations
Citations of this article
45Readers
Mendeley users who have this article in their library.

Abstract

This work develops a rapid and highly sensitive electrochemical sensor for creatinine detection at platinum microelectrode arrays (Pt-MEA). Copper(ii) ions are introduced to form the electroactive creatinine complex, which is then detected at Pt-MEA through a direct reduction reaction. Electrochemical behaviors of the creatinine complex are also explored at Pt macrodisc and microdisc electrodes in comparison with Pt-MEA. At the Pt-MEA, the linear range, sensitivity, and limit of detection of creatinine are determined to be 0.00-5.00 mM, 5401 ± 99 A m−2 M−1, and 0.059 mM (3SB/m), respectively. Notably, the Pt-MEA requires only 10 μL of sample and allows direct measurement of creatinine in synthetic urine with 97.39 ± 4.78% recovery.

Cite

CITATION STYLE

APA

Kaewket, K., & Ngamchuea, K. (2023). Electrochemical detection of creatinine: exploiting copper(ii) complexes at Pt microelectrode arrays. RSC Advances, 13(47), 33210–33220. https://doi.org/10.1039/d3ra06175b