Interlaboratory testing of porcine sera for antibodies to porcine circovirus type 2

17Citations
Citations of this article
18Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A panel of 20 porcine sera was distributed to 5 laboratories across Europe and Canada. Each center was requested to test the sera for the presence of porcine circovirus type 2 antibodies using the routine assays, indirect immunofluorescence assay (IFA) and indirect immunoperoxidase monolayer assay (IPMA), and to determine the titer of each serum. Results from all centers were then compiled and correlated. They demonstrate a wide variation in the titers obtained between laboratories. These differences were dependent on the assay used and the choice of fixative. In general, IPMA gave higher titers than did IFA, and paraformaldehyde gave higher titers than did acetone or ethyl alcohol. This report highlights the need for standardized procedures and biologicals for this virus.

Cite

CITATION STYLE

APA

McNair, I., Marshall, M., McNeilly, F., Botner, A., Ladekjaer-Mikkelsen, A. S., Vincent, I., … Rhodes, C. (2004). Interlaboratory testing of porcine sera for antibodies to porcine circovirus type 2. Journal of Veterinary Diagnostic Investigation, 16(2), 164–166. https://doi.org/10.1177/104063870401600214