Photoacoustic investigation of QCL modulation techniques

2Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

High detection sensitivity and spectral selectivity is important for gas analysers to identify the measured compound and to detect low concentrations. We investigated three different modulation methods - pulse gate modulation, pulse frequency modulation and chopper modulation - for a new pulsed quantum cascade laser based photoacoustic sensor. The spectral selectivity and the detection limit for the three modulation methods are compared by measuring nitric oxide absorption lines and different concentrations. The highest detection sensitivity of 70 ppb was achieved with pulse gate modulation but at the lowest spectral resolution. The highest spectral resolution was achieved with chopper modulation but at the lowest detection sensitivity. It is demonstrated that for the three modulation methods a compromise has to be made between selectivity and sensitivity for each measuring task. © 2010 IOP Publishing Ltd.

Cite

CITATION STYLE

APA

Germer, M., & Wolff, M. (2010). Photoacoustic investigation of QCL modulation techniques. In Journal of Physics: Conference Series (Vol. 214). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/214/1/012067