Calibration of Infra-red CO 2 Gas Analyzers

  • Bate G
  • D'Aoust A
  • Canvin D
N/ACitations
Citations of this article
22Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Precision gas mixing pumps produce CO(2) gas mixtures for the calibration of infra-red CO(2) gas analyzers equivalent in accuracy to the standard CO(2) gas mixtures (+/- 1%) supplied by the National Bureau of Standards, Washington, D. C.The calibration of infra-red gas analyzers by the pressure difference method and by concentration differences did not agree. A factor of 0.709 was necessary to correct the pressure difference method for the effects of pressure broadening with our analyzer.

Cite

CITATION STYLE

APA

Bate, G. C., D’Aoust, A., & Canvin, D. T. (1969). Calibration of Infra-red CO 2 Gas Analyzers. Plant Physiology, 44(8), 1122–1126. https://doi.org/10.1104/pp.44.8.1122