Abstract
Transcatheter aortic valve implantation (TAVI) to manage structural bioprosthetic valve deterioration has been successful in mitigating the risk of a redo cardiac surgery. However, TAVI-inTAVI is a complex intervention, potentially associated with feared complications such as coronary artery obstruction. Coronary obstruction risk is especially high when the previously implanted prosthesis had supra-annular leaflets and/or the distance between the prosthesis and the coronary ostia is short. The BASILICA technique (bioprosthetic or native aortic scallop intentional laceration to prevent iatrogenic coronary artery obstruction) was developed to prevent coronary obstruction during native or valve-in-valve interventions but has now also been considered for TAVI-in-TAVI interventions. Despite its utility, the technique requires a not so widely available toolbox. Herein, we discuss the TAVI-in-TAVI BASILICA technique and how to perform it using more widely available tools, which could spread its use.
Author supplied keywords
Cite
CITATION STYLE
Tagliari, A. P., Saadi, R. P., Medronha, E. F., & Saadi, E. K. (2021, December 1). The use of basilica technique to prevent coronary obstruction in a tavi-tavi procedure. Journal of Clinical Medicine. MDPI. https://doi.org/10.3390/jcm10235534