Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: A survey among anaemic Gambian infants aged 6–12 months

2Citations
Citations of this article
12Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Background In low-income countries, point-of-care photometers are used in the screening and management of anaemia in individuals, but also in the assessment of population iron status when evaluating efficacy of intervention studies or public health interventions. Aims We aimed to evaluate the accuracy of a commonly used photometer, HemoCue-301, in determining haemoglobin concentration among anaemic children aged 6–12 months in a field setting in rural Africa. Methods This report concerns a secondary analysis of data from Gambian infants being screened for an ongoing randomized controlled trial. In those found to be anaemic by HemoCue-301, haemoglobin concentration was measured by Sysmex XN-1500, an automated haematology analyser that was used as a reference. Passing-Bablok regression analysis was used to estimate the regression constant (systematic deviation between two measurement methods that remain consistent across the range of measurements) and proportional bias (systematic deviation between two measurement methods that change in magnitude relative to the value being measured). Results Analysis was based on 227 participants. There was strong evidence of absolute bias among moderately anaemic participants (haemoglobin concentration at 8.0g/dL) (absolute bias: 1.12g/dL; 95% CI: 0.91 to 1.37g/dL; proportional bias: 14.0%; 95% CI: 11.4% to 17.1%) in haemoglobin concentrations measured by HemoCue-301 compared to those measured by Sysmex XN-Series1500. Bias was marginal at haemoglobin concentration of 11.0g/dL (absolute bias: -0.08g/dL; 95% CI: -0.18 to 0.07g/dL; proportional bias: -7.3%; 95% CI: -6.5% to 0.6%). Conclusion Haemoglobin measurements by HemoCue-301 seem substantially biased in participants with haemoglobin less than 8.0g/dL.

Cite

CITATION STYLE

APA

Bah, M., Verhoef, H., Camara, A., Ngom, M. N., Jallow, D., Bajo, K., … Cerami, C. (2024). Comparison of haemoglobin concentration measurements using HemoCue-301 and Sysmex XN-Series 1500: A survey among anaemic Gambian infants aged 6–12 months. PLoS ONE, 19(11). https://doi.org/10.1371/journal.pone.0310577

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free