On the traceably accurate voltage calibration of electrostatic accelerators

43Citations
Citations of this article
24Readers
Mendeley users who have this article in their library.
Get full text

Abstract

We describe in detail a calibration method for the terminal voltage of small accelerators used for ion beam analysis, with the elastic resonance of 16O(α,α)16O at 3038 keV as the intrinsic measurement standard. The beam energy relative to this resonance is determined with a precision around 300 eV and an evaluated reproducibility of 1.0 keV. We show that this method is both robust and convenient, and demonstrate consistency with calibration relative to three other independent methods: using radioactive sources and using the resonant 27Al(p,γ)28Si and non-resonant 16O(p,γ)17F direct capture reactions. We re-evaluate the literature and show that the peak in the cross-section function is at 3038.1 ± 2.3 keV. By comparing the results obtained with 16O(α,α)16O to the other calibration methods we show that this uncertainty can be reduced to 1.3 keV.

Cite

CITATION STYLE

APA

Colaux, J. L., Terwagne, G., & Jeynes, C. (2015). On the traceably accurate voltage calibration of electrostatic accelerators. Nuclear Instruments and Methods in Physics Research, Section B: Beam Interactions with Materials and Atoms, 349, 173–183. https://doi.org/10.1016/j.nimb.2015.02.048