Before implementing one of two new LVEF radionuclide gated ventriculogram (MUGA) systems, the results from 312 consecutive parallel patient studies were evaluated. Each gamma-camera acquisition was simultaneously processed by semi-automatic Medasys Pinnacle and by fully automatic and semiautomatic Philips nuclear medicine computer systems. The Philips systems yielded LVEF results within ±5 LVEF percentage points of the Medasys system in fewer than half of the studies. The remaining values were higher or lower than those from the long-used Medasys system. These differences might have changed cancer patient chemotherapy clinical decisions. As a result, our institution elected not to implement either new system.
Author supplied keywords
Cite
CITATION STYLE
Fair, J. R., Heintz, P. H., & Telepak, R. J. (2009). Evaluation of new data processing algorithms for planar gated ventriculography (MUGA). Journal of Applied Clinical Medical Physics, 10(3), 173–179. https://doi.org/10.1120/jacmp.v10i3.2977