Abstract
Electrospray ionization performs best with volatile buffers. However, generally the best separation performance for capillary electrophoresis (CE) is achieved with non-volatile buffers. Hyphenation of CE with mass spectrometry (MS) utilizing atmospheric pressure photoionization (APPI) enables use of a wider range of separation buffers without compromising detection sensitivity. As APPI is considered to be mass flow sensitive, the use of a larger inner diameter separation capillary (75mm) allows larger volumes to be injected, without decreased separation performance, thus providing improved sensitivity (approx. a factor of 10), compared to the use of a 25 μm capillary. However, nebulizing gas flow and position of capillary tip in the sprayer have to be carefully optimized to prevent excessive band broadening. Further improvement in sensitivity (approx. a factor of 2) was obtained by decreasing the distance between the sprayer and ionization region, indicating that a specially designed CE/APPI-MS interface for low flow rates will be favourable. © 2010 John Wiley & Sons, Ltd.
Cite
CITATION STYLE
Axén, J., Malmström, D., Axelsson, B. O., Petersson, P., & Sjöberg, P. J. R. (2010). Efforts to improve detection sensitivity for capillary electrophoresis coupled to atmospheric pressure photoionization mass spectrometry. Rapid Communications in Mass Spectrometry, 24(9), 1260–1264. https://doi.org/10.1002/rcm.4510