Abstract
Background: Nickel titanium (NiTi) rotary files have drastically altered the treatment protocol in endodontics, allowing for faster and easier preparation and more thorough irrigation of the root canal system. Despite the advantages of the NiTi files, instrument separation still remains a major concern. The aim of this study was to compare the cyclic fatigue resistance of three different single-file NiTi systems after clinical use: WaveOne Gold (WOG, Dentsply Maillefer, Ballaigues, Switzerland), One Curve (OC, Micro Mega, Besancon, France), and Reciproc Blue (RPC Blue, VDW, Munich, Germany). Methods: A total of 120 patients requiring endodontic treatment to first or second mandibular molars, were randomly divided into three groups (patient n = 40) and root canals were prepared with WOG, OC or RPC Blue. Each group consists of 2 subgroups: first use (file n = 10, patient n = 10) and third use of file (file n = 10, patient n = 30). In control group, files were subjected to cyclic fatigue test without use (file n = 10). Cyclic fatigue resistance of files was assessed by groove method. The data was evaluated by two-way ANOVA, Games & Howell post hoc and Weibull reliability analyses (p