Abstract
Purpose: A preliminary study in order to test the accuracy of the IOL Master in detecting axial eye length changes after laser-assisted in situ keratomileusis and to correlate the findings with the theoretical ablation depth. Methods: Pre- and postoperative axial eye length measurements were performed in 10 consecutive eyes that underwent laser-assisted in situ keratomileusis with the Alcon Ladar Vision excimer laser to treat myopic refractive errors ranging from -2.50 to -8.00 diopters of spherical equivalent (mean: -5.23 ± 1.30 D). Results: The preoperative axial eye length measurements ranged from 24.53 mm to 27.78 mm (mean: 25.80 ± 1.01 mm) and from 24.57 mm to 27.57 mm (mean: 25.68 ± 0.93 mm) 1 month after the operation. Preoperative and 1-month postoperative data showed a statistically significant difference (P < 0.05). Conclusions: The IOL Master showed a decrease in the axial eye length measurements that was bigger than the theoretical ablation depth and did not correlate well with it. Such data need to be confirmed by a larger cohort of patients. © 2010 Chalkiadakis et al.
Author supplied keywords
Cite
CITATION STYLE
Chalkiadakis, S. E., Amariotakis, G. A., Parikakis, E. A., & Peponis, V. G. (2010). Axial eye length measurements pre- and postlaser- assisted in situ keratomileusis using the IOL Master: A pilot study. Clinical Ophthalmology, 4(1), 1267–1269. https://doi.org/10.2147/OPTH.S14332