Abstract
Fading is one of the major obstacles to reliable observation in fluorescence microscopy. Using a confocal laser scanning microscope (CLSM) coupled to a computer, we quantitatively measured fading of fluorescence to formulate an equation, evaluated the anti-fading ability of several anti-fading media, and restored the faded images to the original level according to this equation. NIH 3T3 cells were stained with fluorescein isothiocyanate (FITC)-phalloidin, mounted with several commercial and homemade anti-fade media, and observed with CLSM under repeated illumination. With any mounting medium, attenuation of fluorescence intensity at a certain pixel occurred stepwise and the decrease was proportional to the intensity of the previous scan. From these results, we formulated an equation that has three coefficients: anti-fading factor (A), indicating the ability to retard fading; fluorescent intensity at the first scan (EM1); and background fluorescence (B). The fluorescent intensity at a certain point following nth scan is given as EMn = EM1 · A (n-1). This equation was available for restoring faded images to their original states, even after the image had faded to only 60% of its original intensity.
Author supplied keywords
Cite
CITATION STYLE
Ono, M., Murakami, T., Kudo, A., Isshiki, M., Sawada, H., & Segawa, A. (2001). Quantitative comparison of anti-fading mounting media for confocal laser scanning microscopy. Journal of Histochemistry and Cytochemistry, 49(3), 305–311. https://doi.org/10.1177/002215540104900304