Isoelectric PVCs diagnosed with a monitor with simultaneous ECG, and ICD intracardiac electrogram and marker channels

2Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

A 78-year-old woman with a remote history of chronic atrial fibrillation and catheter ablation of the atrioventricular node with a permanent ventricular pacemaker sustained a syncopal episode. Interrogation of her pacemaker revealed electrograms of ventricular fibrillation with a duration of 26 seconds. The patient underwent upgrade to an implantable cardioverter defibrillator (ICD). Postimplant, on telemetry, pauses were seen without pacing artifact, suspicious for oversensing of something by the ICD. Interrogation and threshold testing of the ICD established that both sensing (of her junctional escape rhythm) and pacing thresholds were excellent. A Holter monitor was utilized to evaluate the patient's rhythm and device function. It incorporated simultaneous tracings from multiple surface electrocardiogram leads and both intracardiac electrograms and marker channel recordings from the ICD. The causes for the pauses were found to be isoelectric premature ventricular complexes not seen on the telemetry tracings. © 2009, The Authors.

Author supplied keywords

Cite

CITATION STYLE

APA

Scher, D. L., Man, D. C., Edery, T., & Lu, S. (2009). Isoelectric PVCs diagnosed with a monitor with simultaneous ECG, and ICD intracardiac electrogram and marker channels. PACE - Pacing and Clinical Electrophysiology, 32(1), 137–139. https://doi.org/10.1111/j.1540-8159.2009.02188.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free